Images: Alt
Product Short Description: RXSOL brand highly purified and well balanced Barium Chloride 10% solution chemically plays role during Sulfate Test, to pricipitate as Barium Sulfate crystal of uniform size.
Product Description:

RXSOL brand highly purified and well balanced Barium Chloride 10% solution chemically plays role during Sulfate Test, to pricipitate as Barium Sulfate crystal of uniform size. 

Other Higher Concentration Product ::: Barium Chloride saturated Solution

Testing method of sulfite with barium chloride is based on the precipitation of sulfate ion in a hydrochloric acid medium with barium chloride in such a manner as to form barium sulfate crystals of uniform size.  The glycerol sulfate solution is used to maintain proper crystal size, and control of the stirring and waiting times is necessary to obtain reproducible result.  The resultant turbidity is measured at a wavelength sensitive to turbidity on a photometer.  The blank serves to partially cancel out interferences due to color and suspended matter. 

 

Click here for Testing Procedure  ::: rxmarine.com/SULFATE-TEST-KIT

Product Application:

Water Testing Procedures Manual ( Full Range Lab testing   Methods with standard  composition ratio Of Reagents  ::: Available with us  at 2700 INR or 54 US$ or 207 AED

Product Dose:

Apparatus Required :::  Filter photometer complete with assorted laboratory glassware. 

Chemicals Required

Glycerol Sulfate Solution
Barium Chloride (Natural )  Solution
Sulfate Standard

Procedure of Test


This procedure uses a wavelength of 420 nm and a cell with a light path of approximately 10-mm. First prepare a calibration curve for the photometer using successive dilutions of the sulfate standard to adequately cover the range of sulfate in the samples to be tested.  The dilutions of the standard should be treated in exactly the same manner as that shown below for analysis of the water samples.  Using a 50-ml plastic graduate, add 50 ml of sample to a 125-ml glass flask. Pipet 10 ml of glycerol sulfate solution into the flask and swirl the solution. Pipet 5 ml of distilled water into the flask, swirl the solution and use this solution to set the photometer at zero absorbance. Discard the solution and rinse the flask with distilled water. Into a second 50-ml sample, pipet 10 ml of glycerol sulfate solution and swirl. Adjust the temperature of the sample to 70 to 75F, and pipet 5 ml of neutral barium chloride solution into the same flask. Stir the solution by rotating a rubber paddle between the palms at moderate speed for exactly five seconds. Wait exactly five minutes, and then measure the absorbance on the photometer. 

Calculation of Results  The sulfate concentration in parts per million as SO4  is obtained by reference to the prepared sulfate calibration curve. 

Limitations of Test 

Color or large amounts of suspended matter will interfere with this method,  but some suspended matter may be removed by filtration. Also the presence of color and suspended matter is compensated for to some extent by the blank determination. Silica in excess of 500 ppm will interfere, and in samples containing large amounts of organic material, it may not be possible to precipitate barium sulfate satisfactorily.  

Product Note:
Product Technical Specification:
Remarks:
prod_useful_area:
Product Supply Location:
stock-clearance.png