Sulphite (Sulfite) Test Kit HANDY EASY METHOD

Size:

Color:

Product Short Description: The determination of sulfite usually is made only on boiler waters or on waters that have been treated with catalyzed sodium sulfite for corrosion prevention.
Product Technical Specification:

Interfering substance

Interference level Metals Some metals, especially copper, catalyze the oxidation of sulfite to sulfate. Immediately add  TK - 14 ::: RXSOL-62-5503-003 powder for each liter of sample during sample collection to prevent the interference.

Nitrite Reacts with sulfite and causes low results.

Organic compounds Oxidizable organic compounds can cause high results.

Oxidizable compounds Cause high results.

Sulfide Causes high results. 

Remarks:

Samples must be analyzed immediately after collection and cannot be preserved for later analysis. 

Let the sample temperature decrease to 50 °C (122 °F) or less before analysis. 

Sulfite reacts quickly with oxygen in the air. Shaking or swirling the sample causes low results. Prevent agitation of the sample during the procedure. 

Review safety chemical handling procedure. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Follow the environmental safety, health safety staff for your facility and/or local regulatory agencies for further disposal information.

Images: Alt



Boiler Water Test Kit 311

Size:

Color:

Product Short Description: Boiler Water Test Kit 311
Product Technical Specification:
Remarks:
Images: Alt



Alkalinity Test Kit p-m Total

Size:

Color:

Product Short Description: Alkalinity TEST FOR BOILER and COOLING WATER
Product Technical Specification:


OFFER :::---
Do not forget to ask JUMBO PACK OFFER  (  TK - extra pack )

Ask for  FREE TK extra pack

Remarks:

Avoid contact with EYE / Hand and other CHEMICALS and Gases.
 

For determination of Hydroxide alkalinity only, barium chloride is added prior to titration to precipitate the carbonate ions.  Titration is then taken only to the phenolphthalein end point.  A pH meter may be used instead of the indicators to determine the end points of the titration.

Images: Alt
Get MSDS
Get Technical Bulletin
Connect On Whats App



Nitrite Test Kit

Size:

Color:

Product Short Description: Nitrites are useful as corrosion inhibitors, preservatives, pigments, and in manufacturing many organic preservative chemicals. A Maximum Contaminant Level of 1 mg/L has been established by the USEPA for nitrite-nitrogen in drinking water.
Product Technical Specification:

OFFER :::---
Do not forget to ask JUMBO PACK OFFER  (  TK - extra pack )

BUY 2 - GET 2   FREE TK extra pack

Remarks:
Images: Alt
Get MSDS
Get Technical Bulletin
Connect On Whats App



Chloride Test Kit

Size:

Color:

Product Short Description: Determination of Chloride ppm value
Product Technical Specification:

OFFER :::---
Do not forget to ask JUMBO PACK OFFER  (  TK - extra pack )

BUY 2 - GET 2   FREE TK extra pack

Remarks:


REAGENT & APPARATUS:
Reagent:  RXSOL TK 1 , TK 2 , TK 3 , TK4 . Apparatus : TEST TUBE , Titrator or DROPPER.

VIDEO Help :  
https://www.youtube.com/watch?v=p_IgKoO49iE

Images: Alt
Get MSDS
Get Technical Bulletin
Connect On Whats App



Hydrazine Test Kit Colorimetric Procedure

Size:

Color:

Product Short Description: Distinct yellow colour test
Product Technical Specification:

Hydazine Dose Level

Thumb rule for Hydarzine dose is, Low pressure boiler required high dose of Hydrazine ( 0.1 to 0.2 ppm level of hydrazine ) While for High pressure boiler low dose of hydrazine required (0.05 to 0.1ppm value)
 
Recommended dose of Hydrazine :

Low Pressure Boiler ( 0 to 40 bar )

0.10 to 0.2  ppm level  

Medium Pressure Boiler ( 41 to 60 bar )

 0.1 to 0.15 ppm level  

High Pressure Boiler ( more then 60 bar) 

0.05 to 0.10 ppm level

Remarks:

Before Sampling all PRECAUTION should be taken :::

Toxic effects of Hydrazine ( which is collected in sample ) is a severe skin and mucous membrane irritant in humans; in animals.In humans, the vapor is immediately irritating to the nose and throat and causes dizziness and nausea; itching, burning, and swelling of the eyes develop over a period of several hours. Severe exposures of the eyes to the vapors causes temporary blindness lasting for about 24 hours. Recurrent exposure to hydrazine hydrate has been reported to cause contact dermatitis of the hands without systemic intoxication. In humans, hydrazine is absorbed through the skin, by inhalation, and orally; systemic effects include weight loss, weakness, vomiting, excited behavior, and convulsions; the chief histologic findings are fatty degeneration of the liver and nephritis. 

Images: Alt
Get MSDS
Get Technical Bulletin
Connect On Whats App



Cooling Water Test kit RXSOL 309

Size:

Color:

Product Short Description: REAGENT:-NITRITE, CHLORIDE , pH, ALKALINITY
Product Technical Specification:
Remarks:
Images: Alt
Get MSDS
Get Technical Bulletin
Connect On Whats App



Boiler Water Test Kit 310

Size:

Color:

Product Short Description: Boiler Water Test Kit 310
Product Technical Specification:
Remarks:
Images: Alt
Get MSDS
Get Technical Bulletin
Connect On Whats App



Test Tube Stand for 50 Ml Stainless Steel

Size:

Color:

Product Short Description: Anti Rust chemical resistance Stainless Steel stand
Product Technical Specification:
Remarks:
Images: Alt
stock-clearance.png